The purpose of the PRESSURE-1723 ILC/PT is to give organizations an opportunity to demonstrate their proficiency (e.g., use of appropriate equipment, procedures, ability to follow instructions) in the calibration of a Digital Pressure Gauge. This particular cage can use both types of media to drive the internal mechanisms.
Participants are asked to make 9 pressure measurements from 0 to 10,000 psi in ascending and descending set points. The resolution of the digital artifact is 1 psi. Pressure medium - use only clean, dry non-corrosive gases (pneumatic). The manufacturer specification for artifact accuracy is 0.1% FS (FS = 10,000 psi).
Participants are expected to be familiar with and have appropriate lab equipment and measurement uncertainties for calibrating similar devices.
Artifact in kit: Additel, ADT680, 0 to 10,000 psi